Haneveld, J. and Lammerink, T.S.J. and de Boer, M.J. and Sanders, R.G.P. and Mehendale, A. and Lötters, J.C. and Dijkstra, M.A. and Wiegerink, R.J. (2010) Modeling, design, fabrication and characterization of a micro Coriolis mass flow sensor. Journal of micromechanics and microengineering, 20 (6). 125001. ISSN 0960-1317 *** ISI Impact 2,276 ***

Full text available as:

PDF
- Univ. of Twente only - Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader
1940 Kb

Official URL: http://dx.doi.org/10.1088/0960-1317/20/12/125001

Exported to Metis

Abstract

This paper discusses the modeling, design and realization of micromachined Coriolis mass flow sensors. A lumped element model is used to analyze and predict the sensor performance. The model is used to design a sensor for a flow range of 0–1.2 g h−1 with a maximum pressure
drop of 1 bar. The sensor was realized using semi-circular channels just beneath the surface of a silicon wafer. The channels have thin silicon nitride walls to minimize the channel mass with respect to the mass of the moving fluid. Special comb-shaped electrodes are integrated on the
channels for capacitive readout of the extremely small Coriolis displacements. The comb-shaped electrode design eliminates the need for multiple metal layers and sacrificial layer etching methods. Furthermore, it prevents squeezed film damping due to a thin layer of air between the capacitor electrodes. As a result, the sensor operates at atmospheric pressure with a quality factor in the order of 40 and does not require vacuum packaging like other micro
Coriolis flow sensors. Measurement results using water, ethanol, white gas and argon are presented, showing that the sensor measures true mass flow. The measurement error is
currently in the order of 1% of the full scale of 1.2 g h−1.

Item Type:Article
Research Group:EWI-TST: Transducers Science and Technology, EWI-CE: Control Engineering
Research Program:MESA-General
Research Project:Micro Coriolis flow Enabling
ID Code:19119
Status:Published
Deposited On:23 December 2010
Refereed:Yes
International:Yes
ISI Impact Factor:2,276